Error Diagnostics of Eccentric Loadings of Single Point Load Cells
Show PDF version
    - If the load is applied eccentrically (not in the centre of the load cell), an error
        can appear. This error is maximum ± 0,0000085 % / cm.
        
			
<
		/div>
	
    
- Example: Platform size 80 x 60 cm.
        If a load is applied and its centre of gravity is in a radius of 25 cm from the load
        cells centre (r = 25 cm), the indicated signal can show an error of
        25 cm x 0.0000085 % / cm = 0.00002125 %.
        In clearer values:
        
	 
    - To reduce such an error, the load should be applied – if possible – in the centre
        of the platform (directly over load cell), or well-proportioned all over the platform.
 
    - Extreme eccentric loadings should be avoided.
        They not only create errors in the readings, but can also damage the load cell
        due to the leverage effect.
        This can be seen when the zero signal changes, so the tare value.